The Newcastle University research output collection, currently available on ePrints, will shortly be moving to a new open repository platform, Figshare. To prepare for the data migration we have paused adding new content to ePrints, and will resume once the new repository is launched. During this time you will continue to have access to ePrints (but no new content will appear). We will share updates here when available.

Automated MR-only radiotherapy outperforms CT-based radiotherapy and decreases hands-on time for head-and-neck cancer treatment

Lookup NU author(s): JJ Wyatt, Dr Rachel Pearson

Downloads


Licence

This work is licensed under a Creative Commons Attribution 4.0 International License (CC BY 4.0).


Abstract

© 2026 The Authors. Published by Elsevier B.V. on behalf of European Society for Radiotherapy and Oncology. This is an open access article under the CC BY license. http://creativecommons.org/licenses/by/4.0/Introduction: Cancer incidence is expected to increase in Europe by 18% in eighteen years. To account for the increasing patient numbers, the workload per patient needs to be reduced. One step towards future-proof radiotherapy is automated MR-only radiotherapy as it could eliminate the need for (i) a planning CT and (ii) for manual organ at risk (OAR) delineations. The aim of this study was to evaluate the feasibility of an automated MR-only workflow for head-and-neck radiotherapy. Method: Automated MR-only radiotherapy consisted of a Zero-Echo-Time-based synthetic CT for dose calculations and automated T2w-based OAR delineations. Automated MR-Only RT was compared to the clinical workflow consisting of CT-based dose calculation and CT-based OAR delineations. Both approaches were benchmarked to a gold standard consisting of the planning CT for dose calculations and manual delineations on the T2w MR scan. Dice similarity coefficients (DSC), 95% Hausdorff distances and absolute DVH metrics were compared between the clinical and MR-only workflow using a linear mixed-effect model. A p-value < 0.05 was deemed significant. Results: Seventeen head-and-neck cancer patients were included. The automated MR-only delineations were more accurate compared to the clinical CT delineations (DSC of 0.79 vs. 0.67; 95% Hausdorff distance 4.0 vs 5.8 mm (p-values < 0.001)). The average dose calculation errors of the automated MR only RT were smaller than the clinical workflow (+0.34 Gy vs. −1.39 (p-value < 0.01)). Discussion: The automated MR-only head-and-neck radiotherapy workflow was more accurate than the standard CT based clinical workflow, demonstrating the feasibility of automated MR-only RT to decrease the workload for head-and-neck RT treatment preparation.


Publication metadata

Author(s): Lauwers I, Capala M, Kaushik S, Rusko L, Cozzini C, Szabo E, Kekesi A, Deak-Karancsi B, Wyatt J, Pearson R, Verduijn G, Wiesinger F, Hernandez-Tamames J, Petit S

Publication type: Article

Publication status: Published

Journal: Clinical and Translational Radiation Oncology

Year: 2026

Volume: 58

Print publication date: 01/05/2026

Online publication date: 13/02/2026

Acceptance date: 07/02/2026

Date deposited: 21/04/2026

ISSN (electronic): 2405-6308

Publisher: Elsevier Ireland Ltd

URL: https://doi.org/10.1016/j.ctro.2026.101122

DOI: 10.1016/j.ctro.2026.101122


Altmetrics

Altmetrics provided by Altmetric


Funding

Funder referenceFunder name
EIT Health
Deep MR-only Radiation Therapy activity (project numbers: 19037, 20648, 210995)
European Union Horizon 2020 Research and innovation program