The Newcastle University research output collection, currently available on ePrints, will shortly be moving to a new open repository platform, Figshare. To prepare for the data migration we have paused adding new content to ePrints, and will resume once the new repository is launched. During this time you will continue to have access to ePrints (but no new content will appear). We will share updates here when available.

Phase-contrast and magnification radiography at diagnostic X-ray energies using a pseudo-microfocus X-ray source

Lookup NU author(s): Dr Colin Kotre, Dr Kevin Robson

Downloads

Full text for this publication is not currently held within this repository. Alternative links are provided below where available.


Abstract

© 2014 The Authors. Objective: To investigate the use of conventional diagnostic X-ray tubes for applications in which specialist microfocus sources are normally required. Methods: A conventional diagnostic X-ray tube was used in conjunction with a range of apertures to investigate improvements in spatial resolution using a line-pairs test object. Phase-contrast effects were investigated by varying source-to-object and object-to-receptor distances using a 2-French catheter as a clinically realistic test object. Results: For magnification radiography using a computed radiography receptor and conventional X-ray tube with a 1-mm nominal focus size, the limiting spatial resolution was improved from 3.55 line-pairs per millimetre, for a conventional contact image, to 5.6 line-pairs per millimetre, for a 32 magnified view with a 250-mm aperture. For inline phase-contrast radiography, phase contrast enhancement of a 2-French catheter was demonstrated, and the expected trends with variations in source-toobject and object-to-receptor distances were found. Images of a neonatal phantom demonstrated a subtle improvement in visibility of a superimposed 1-French catheter simulating a percutaneously inserted central catheter for no increase in patient radiation dose. Conclusion: Spatial resolution improvement and visible phase contrast can be produced in clinically relevant objects using a pseudo-microfocus geometry at X-ray energies in the normal diagnostic range, using conventional diagnostic X-ray tubes and image receptors. The disadvantages of the proposal are the large distances required to produce phase contrast and limitations imposed by the resulting tube loading. Advances in knowledge: It is possible to use conventional diagnostic X-ray equipment in applications that normally require microfocus X-ray sources. This presents some possibilities for clinical applications.


Publication metadata

Author(s): Kotre CJ, Robson KJ

Publication type: Article

Publication status: Published

Journal: British Journal of Radiology

Year: 2014

Volume: 87

Issue: 1039

Online publication date: 29/05/2014

Acceptance date: 28/04/2014

ISSN (print): 0007-1285

ISSN (electronic): 1748-880X

Publisher: British Institute of Radiology

URL: https://doi.org/10.1259/bjr.20130734

DOI: 10.1259/bjr.20130734

PubMed id: 24779409