The Newcastle University research output collection, currently available on ePrints, will shortly be moving to a new open repository platform, Figshare. To prepare for the data migration we have paused adding new content to ePrints, and will resume once the new repository is launched. During this time you will continue to have access to ePrints (but no new content will appear). We will share updates here when available.

A user-operated audiometry method based on the maximum likelihood principle and the two-alternative forced-choice paradigm.

Lookup NU author(s): Dr Christian Brandt

Downloads

Full text for this publication is not currently held within this repository. Alternative links are provided below where available.


Abstract

Objective: To create a user-operated pure-tone audiometry method based on the method of maximum likelihood (MML) and the two-alternative forced-choice (2AFC) paradigm with high test-retest reliability without the need of an external operator and with minimal influence of subjects’ fluctuating response criteria. User-operated audiometry was developed as an alternative to traditional audiometry for research purposes among musicians. Design: Test-retest reliability of the user-operated audiometry system was evaluated and the user-operated audiometry system was compared with traditional audiometry. Study sample: Test-retest reliability of user-operated 2AFC audiometry was tested with 38 naïve listeners. User-operated 2AFC audiometry was compared to traditional audiometry in 41 subjects. Results: The repeatability of user-operated 2AFC audiometry was comparable to traditional audiometry with standard deviation of differences from 3.9 dB to 5.2 dB in the frequency range of 250–8000 Hz. User-operated 2AFC audiometry gave thresholds 1–2 dB lower at most frequencies compared to traditional audiometry. Conclusions: User-operated 2AFC audiometry does not require specific operating skills and the repeatability is acceptable and similar to traditional audiometry. User operated 2AFC audiometry is a reliable alternative to traditional audiometry.


Publication metadata

Author(s): Schmidt JH, Brandt C, Pedersen ER, Christensen-Dalsgaard J, Andersen T, Poulsen T, Bælum J

Publication type: Article

Publication status: Published

Journal: International journal of audiology

Year: 2014

Volume: 53

Issue: 6

Pages: 383-391

Print publication date: 01/06/2014

Online publication date: 01/02/2014

Acceptance date: 23/12/2013

ISSN (print): 1499-2027

ISSN (electronic): 1708-8186

Publisher: Informa Healthcare

URL: http://dx.doi.org/10.3109/14992027.2013.879339

DOI: 10.3109/14992027.2013.879339


Altmetrics

Altmetrics provided by Altmetric


Share